Assessment of efficacy of ultraviolet chamber in disinfecting dental instruments

Authors

  • Rahul G. Naik Senior Lecturer, Department of Public Health Dentistry,A.C.P.M. Dental College, Dhule, Maharashtra, India. Author
  • Arun S. Dodamani Principal, Professor & Head, Department of Public Health Dentistry,A.C.P.M. Dental College, Dhule, Maharashtra, India. Author
  • Prashanthkumar Vishwakarma Reader, Department of Public Health Dentistry,A.C.P.M. Dental College, Dhule, Maharashtra, India. Author
  • Mahesh R. Khairnar Assistant Professor, Department of Public Health Dentistry, Bharati Vidyapeeth Deemed University Dental College and Hospital, Sangli, Maharashtra, India. Author
  • Harish C. Jadhav Senior Lecturer, Department of Public Health Dentistry,A.C.P.M. Dental College, Dhule, Maharashtra, India. Author
  • Manjiri A. Deshmukh Senior Lecturer, Department of Public Health Dentistry,Swargiya Dadasaheb Kalmegh Smruti Dental College & Hospital, Nagpur, Maharashtra, India. Author

DOI:

https://doi.org/10.5530/5rqf4d25

Keywords:

Colony Forming Units (CFU), Disinfection, Microorganisms, UV Chamber

Abstract

Cross-contamination has been a major concern for both the practitioner and the patient in recent years. Various sterilization equipment's throw the challenge to choose the best available in the market. Hence the present study was conducted to assess the efficacy of UV disinfection chamber.The present study was an in vitro study conducted using contaminated dental instruments. 
Autoclaved dental Instruments were used to obtain the swab from carious lesion. These specimens were then inoculated on blood agar and incubated at 37 C for 24 hrs and colony forming units were determined. Thereafter, these instruments were kept in Ultraviolet (UV) Chamber for 5 minutes and again they were inoculated on blood agar to determine colony forming units. Same experiment was performed for keeping the instruments in UV cabinet for different time periods i.e. 10 minutes, 15 minutes, 20 minutes, 25 minutes, 30 minutes, 45 minutes and 60 minutes. and no. of colony forming units (CFUs) were again determined before and after keeping the instruments in UV cabinet . The values were subjected to descriptive statistical analysis. UVDisinfection Chamber is most effective when instruments are kept in UV chamber for a period of 45-60 minutes. Thus UV Disinfection chambers can be used in dental offices as well as dental outreach program as an effective tool for disinfection of instruments. 

 

 

 

 

Author Biographies

  • Rahul G. Naik, Senior Lecturer, Department of Public Health Dentistry,A.C.P.M. Dental College, Dhule, Maharashtra, India.

     

     
  • Arun S. Dodamani, Principal, Professor & Head, Department of Public Health Dentistry,A.C.P.M. Dental College, Dhule, Maharashtra, India.

     

     
  • Prashanthkumar Vishwakarma, Reader, Department of Public Health Dentistry,A.C.P.M. Dental College, Dhule, Maharashtra, India.

     

     
  • Mahesh R. Khairnar, Assistant Professor, Department of Public Health Dentistry, Bharati Vidyapeeth Deemed University Dental College and Hospital, Sangli, Maharashtra, India.

     

     
  • Harish C. Jadhav, Senior Lecturer, Department of Public Health Dentistry,A.C.P.M. Dental College, Dhule, Maharashtra, India.

     

     
  • Manjiri A. Deshmukh, Senior Lecturer, Department of Public Health Dentistry,Swargiya Dadasaheb Kalmegh Smruti Dental College & Hospital, Nagpur, Maharashtra, India.

     

     

Downloads

Published

2016-12-30

Issue

Section

Articles

How to Cite

Assessment of efficacy of ultraviolet chamber in disinfecting dental instruments. (2016). Asian Journal of Pharmaceutical and Health Sciences, 6(4), 1586-1589. https://doi.org/10.5530/5rqf4d25